Flammability testers are indispensable tools in industries where fire safety is a primary concern. By subjecting materials and products to controlled ignition conditions, these testers help manufacturers, regulators, and consumers make informed decisions about product safety and compliance with industry standards. read more
Flammability Tester LFLC-A14 is a comprehensive solution for assessing fire resistance of plastic and non-metallic materials. It has gas pressure of 0.1Mpa for effective functioning of the tester. Designed with integral type inlet ignition system to ensure safe operation. Multi-functional flame measure gauge provides accurate flame characterization. High precision digital indicator delivers reliable measurements.
Blast Burner Standard | IEC60695-11-3/IEC60695-11-4/10, UL94, inner diameter:Φ9.5±0.3mm,length:100 mm, made of brass |
Gas Pressure | 0.1 Mpa |
Gas Flow | 50W:105 ± 5ml/min or 18 ± 1mm;500W:965±30ml/min or 117 ± 3 mm, flow meter adjustment |
U type pressure gauge back pressure water column height | 50 W:10 mm, 500 W: 125 ± 5 mm |
Test Angle | 50 W: 0°, 45°; 500 W: 20 ° |
Burning Time | 0-9999×0.1 S (Can be preset) |
Residual Flame | 0-9999×0.1S (Can be displayed) |
Test Flame Calibration | Temperature measuring copper head:50W:Ф5.5 mm ± 0.01 mm, weight before drilling: 10 ± 0.05 g |
Thermocouple Temperature Resistance | 1100° |
Distance between the bunsen burner port to temperature measuring copper block | 50 W: 10 ± 1 mm; 500 W: 55 ± 1 mm |
Flame Height |
50W: 20 ± 2 mm (there is no internal and external flame); 500W: Inner flame: 40 ± 2mm,Outer flame:125 ± 10 mm |
Temperature Rise Time | From 100 ± 2° to 700°± 3° 50W: 44 ± 2S |
Test Chamber | > 0.75 cube, black in wall, with exhaust fan |
Power Supply | 220 V/50 Hz |
Power | 50W and 500W double power switching |
Control Mode Operation | PLC control, 7-inch color touch screen operation, with infrared remote control |
Dimension | 1150 × 600 × 1260 mm |
Packaging Dimension | 1210 × 660 × 1420 mm |
Weight | 165 Kg |
Flammability Tester is used across electronics manufacturing, automotives, construction sector, aerospace, etc.
Available Range :
Flammability Tester LFLC-A10 : LCD touch screen display Flammability Tester LFLC- A11 : PLC touch screen display Flammability Tester LFLC-A12 : PLC touch screen display Flammability Tester LFLC-A13 : ± 0.5 ℃ Flammability Tester LFLC-A14 : IEC60695-11-3/IEC60695-11-4/10, UL94, inner diameter:Φ9.5±0.3mm,length:100 mm, made of brass